VAMPIRES Polarization Calibration

❓ TODO: Should we Gen2 archive these cals?

When a PDI run is done on VAMPIRES an associated set of Mueller matrix calibration data should be taken using the AO188 flat lamp. This ensures accurate correction of the instrumental polarization and crosstalk.

To perform PDI calibrations, VAMPIRES must be behind AO188 and an operator or day-crew member must be at the summit to enable the AO188 calibration source. The total time to take calibration data for all filters is 2.5 - 3 hours.

🤝 FastPDI

These calibrations are valuable for FastPDI, too. Consider coordinating with Kudo-san for FastPDI or attempting yourself, since you can archive FastPDI in the background of VAMPIRES with HWP sync without too much hassle.

Setup

Ahead of time decide which HWP angles and image rotator angles you want to test. For now, these must be hand-written inside vControl- talk to Miles or Barnaby for help. Set up the AO188 flat lamp at a higher amplitude, typically 2.4 amps. Test the exposure settings- you'll want to keep the exposure time between 10 and 100 ms.

Flats

Take a series of PDI flats by setting the HWP to continuous rotation and executing a HWP series combine the output cubes to average the effects of instrumental polarization in the flat frames. After this is done you can stop the continuous rotation of the HWP.

Darks

Take darks for the exposure settings found above, can be done after the cal source is removed.

Procedure

1. Insert the polarizer

Insert the linear polarizer using the WPU on garde

2. Run automated procedure

There is an automated function Run AO188 HWP Cal inside vControl which iterates through image rotator and HWP angle combinations.

Screen Shot 2023-02-13 at 01 05 09

You will have to manually execute it for each filter you want to test. You set the data type to "FLAT". You will be prompted to make sure the polarizer is in and you can see the estimated time without overheads to image each rotator and HWP position with the current exposure settings.

⛔️ Abort

If you need to abort, press "abort" and all loops will stop. If you start the cal again, it will start from the beginning of all the image rotator and HWP angle combinations.