The hMRI demo dataset is used and referred to across the whole toolbox, for tutorial and illustration purposes. A description of the data, acquisition parameters and experiment is provided below and in the Data in Brief article: Callaghan2019.
DOWNLOAD
DESCRIPTION OF THE EXPERIMENT
ACQUISITION PARAMETERS
TEMPORARY NOTE: the dataset and protocol examples (detailed list of sequences and acquisition parameters) are currently available from the hMRI toolbox home page, section hMRI Example Data and MRI Protocols
.
- Demo dataset (ZIP archive).
- Corresponding protocol
hmri_sample_dataset_protocol_800um_64ch.pdf
(PDF printout) - Other MPM protocol examples, including Siemens and Philips scanners.
An example multiparameter mapping (MPM) dataset was acquired on a whole body 3T Prisma system (Siemens Healthcare, Erlangen, Germany) to illustrate the hMRI-toolbox features and provide the users with a full MPM dataset used in the tutorial and examples (see for example the step-by-step tutorial for map creation and the toolbox customization examples). The study was approved by the local ethics committee and informed written consent was obtained from the participant prior to scanning.
Rapid calibration data were acquired at the outset of each session to map and consequently correct for inhomogeneities in the B1 transmit field (Lutti2010, Lutti2012). Eleven spin-echo and stimulated-echo pairs were acquired with the nominal flip angles varying from 115° to 65° in 5° decrements.
These were followed by acquisition of 3D multi-echo RF-spoiled gradient echo acquisitions with predominantly PD, T1 or MT weighting similar to the MPM protocol described in Weiskopf2013. The flip angle was 6° for the PDw and MTw images and 21° for the T1w images. MT-weighting was achieved through the application of a Gaussian RF pulse 2 kHz off-resonance with 4 ms duration and a nominal flip angle of 220°. The data were acquired with whole-brain coverage at an isotropic resolution of 800 µm using a FoV of 256 mm head-foot, 224 mm anterior-posterior (AP) and 179 mm right-left (RL). Gradient echoes were acquired with alternating readout gradient polarity at eight equidistant echo times ranging from 2.30 to 18.40 ms in steps of 2.30 ms using a readout bandwidth of 488Hz/pixel. Only six echoes were acquired for the MTw acquisition in order to maintain a TR of 25 ms for all RF-spoiled gradient echo volumes. To accelerate the data acquisition, partially parallel imaging using the GRAPPA algorithm was employed in each phase-encoded direction (AP and RL) with forty reference lines and a speed up factor of two. The acquisition time for each RF-spoiled gradient echo volume was 7'08''. The data were acquired using the body coil for signal transmission and a 64 channel head and neck array for signal reception. Each dataset was acquired with a 30° rotation of the sagittal plane so that any eye-related motion artifact propagated to the neck/inferior cerebellum rather than the cortex.
Prior to the acquisition of each RF-spoiled gradient echo volume, two additional 3D rapid, unaccelerated, low resolution (8 mm isotropic, same FoV) volumes were acquired one with the 64 channel head and neck array coil and the other with the body coil. A single echo, with a TE of 2.20 ms, was acquired in each case using a 6° flip angle and a TR of 6.00 ms. The acquisition time for each of these calibration volumes was 5.90 s.
These data were used to correct for the position-specific sensitivity modulation of the array coil by approximately removing the receive field sensitivity bias. To test the performance of this correction, the participant performed a yaw rotation (i.e. about the z-axis) within the confines of the 64 channel coil prior to the acquisition of the MT-weighted dataset before returning to the original position centered within the head coil for the PD-weighted and T1-weighted acquisitions.
The ZIP archive contains the following data:
-
mfc_seste_b1map_v1e_0004
: B1 mapping data (3D EPI SE/SESTE protocol) -
gre_field_mapping_1acq_rl_0005
: B0 mapping data (2 magnitude images with different TE) -
gre_field_mapping_1acq_rl_0006
: B0 mapping data (presubtracted phase image) -
mfc_smaps_v1a_Array_0007
: RF sensitivity mapping data (64 channel head coil) -
mfc_smaps_v1a_QBC_0008
: RF sensitivity mapping data (body coil) -
pdw_mfc_3dflash_v1i_R4_0009
: 8 echoes with PD-weighting
Large yaw rotation of the volunteer's head (i.e. about the z-axis) within the confines of the 64 channel coil.
-
mfc_smaps_v1a_Array_0010
: RF sensitivity mapping data (64 channel head coil) -
mfc_smaps_v1a_QBC_0011
: RF sensitivity mapping data (body coil) -
mtw_mfc_3dflash_v1i_R4_0012
: 6 echoes with MT-weighting
Back to initial (straight) position.
-
mfc_smaps_v1a_Array_0013
: RF sensitivity mapping data (64 channel head coil) -
mfc_smaps_v1a_QBC_0014
: RF sensitivity mapping data (body coil) -
t1w_mfc_3dflash_v1i_R4_0015
: 8 echoes with T1-weighting
The full protocol is available and can be downloaded as a PDF printout.